AgaMatrix Presto - Troubleshooting Meter Operation Issues; Situation 1: Meter Not Entering Test Mode

AgaMatrix Presto
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
have been removed too early.
[3] If this message appears after the last progress bar instead
of a result, it may indicate that you applied more blood after
testing began.
Error 2:
The meter has detected an irregularity with the sample.
[1] The test strip may be partially filled; you should check the
window of the test strip to confirm that it is full.
[2] The sample may not be blood or control solution.
Error 3:
The meter has detected that the test strip is in poor
condition.
[1] The test strip may have been improperly stored (e.g. hot,
humid conditions) or may be expired.
[2] The test strip may have been mishandled by vigorous
bending or shaking.
[3] Parts of the test strip may have become covered in grease,
oil, or lotion.
[4] The test strip port connector may be dirty.
Error 4:
The meter was unable to produce a result or unusual
test strip problems have occurred that may be related to
extreme conditions.