51
[1] This error may be caused by extremely high glucose. If you
are experiencing symptoms of hyperglycemia, contact your
healthcare professional immediately. If this error code persists
on retesting, consult your healthcare professional.
[2] Retest where it is closer to a room temperature of 70 °F to
75 °F (21 °C to 24 °C).
Error 5:
The meter was unable to produce a reliable result.
[1] This may be caused by a non-blood sample, or a
combination of high glucose and other medical conditions.
[2] If this error code persists on retesting, consult your
healthcare professional.
Error 6:
The test strip has taken too long to generate a signal.
[1] This may be caused by a combination of cold operating
temperature and high hematocrit levels. Retest in a warmer
location.
[2] If this error code persists on retesting, consult your
healthcare professional.
Error 7 with Battery Icon:
Replace batteries immediately before testing.