Recommended protective
distance:
Conducted
high frequency
disturbance
variables in
accordance
with
IEC 61000-4-6
3 V
eff
150 kHz to
80 MHz
3 V
eff
d = 1.2
Radiated high
frequency
disturbance
variables in
accordance
with
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = 1.2 80 MHz to 800
MHz
d = 2.3 800 MHz to 2.5
GHz
With P as the rated power of
the transmitter in watts (W) in
accordance with the
manufacturer information on
the transmitter and d as the
recommended protective
distance in metres (m).
The field strength of stationary
radio transmitters is lower than
the level of the agreement at all
frequencies in accordance with
an on-site investigation.
Disruptions are possible near
devices that carry the following
symbol:
The field strength of stationary transmitters, such as base stations of radio
telephones, mobile broadcasts for rural areas, amateur stations, and AM and
FM radio transmitters, cannot be precisely predetermined theoretically. An
82 | CR 10-X, CR 12-X, CR 15-X | Remarks for HF-emission and immunity