What to do if AGFA Measuring Instruments has image plate transport problems?
- LLisa HarveyAug 3, 2025
If you encounter image plate transport problems with your AGFA Measuring Instruments, remove the cassette, press the confirm button, and then perform the following checks: 1. Inspect the cassette for any signs of damage. 2. Open the cassette and check the shutter for damages. 3. Verify if the image plate is bent by placing it on a flat surface. The entire plate must touch the surface. If there's a gap, like a lifted edge or sharp bend, the image plate needs to be replaced.