Behavior in Case of Power Failure
Note: The description below is only applicable if an uninterruptible
power supply (UPS) is put into the CR system configuration.
In case of a power failure, the system is still connected to the UPS. Two
situations are possible:
Power failure after cassette insertion and before identification with the NX
workstation. The digitizer pushes the image plate back in the cassette
without scanning and releases the cassette. After the power supply
returns, the cassette must be inserted into the digitizer and identified
again to read out the image.
Power failure after identification with the NX workstation. The image
plate is scanned and erased as usual. The scan cycle finishes when the
cassette is released. If the power supply is still not available, the digitizer
will refuse scanning other cassettes.
70 | CR 10-X, CR 12-X, CR 15-X | Operating CR 10-X, CR 12-X and CR 15-X
2491E EN 20150409 1543