Details The following error message is displayed in the Digitizer
Remote Display:
Do not press the cassette release button before end of cycle.
Please clamp the cassette again by pushing it towards the
digitizer. Restart the digitizer.
Cause You pressed the cassette release button before the end of the
cycle.
Solution Do not press the cassette release button before the end of the
cycle. If you did, clamp the cassette again by pushing it
towards the digitizer and restart the digitizer.