CAUTION:
Do not pour AGFA CR Phosphor Plate Cleaner directly on
the image plate.
b) Leave the image plate surface to dry for about 10 minutes to allow the
solvents to evaporate.
c) Once the image plate surface is dry (after about 10 minutes), check
once again for particles of material and other impurities before placing
the image plate into the cassette.
6. Put the image plate back into the cassette.
Verify that the white phosphor side (1) is oriented to the tube side of the
cassette (2) and that the shutter does not scratch the image plate.
1. White phosphor side
2. Tube side of the cassette
Make sure to slide in the image plate carefully. Do not let the
plate fall down vertically into the cassette! This causes
delaminations and damages of the phosphor.
Note: Make certain you put the correct image plate into the
correct cassette. The label on the image plate must
correspond to the label on the cassette.
48 | AGFA CR Plates and Cassettes | Using the Agfa CR Plates and Cassettes
2492F EN 20180417 1049