Chapter 3 Features and Functions
Frequency Sweep
3
Marker Frequency
If desired, you can set the frequency at which the signal on the front-
panel Sync connector goes to a logic low during the sweep. The Sync
signal always goes from low to high at the beginning of the sweep.
Marker frequency: 1 Hz to 30 MHz (limited to 200 kHz for ramp).
The default is 500 Hz.
When the sweep mode is enabled, the marker frequency must be
between the specified start frequency and stop frequency. If you
attempt to set the marker frequency to a frequency not in this range,
the waveform generator will automatically
set the marker frequency
equal to the start frequency or stop frequency
(whichever is closer).
The Sync enable setting is overridden by enabling the marker used
with the sweep mode (see page 107). Therefore, when
the marker is
enabled (and the sweep mode is also enabled
), the Sync setting is
ignored.
Front-Panel Operation: After enabling a sweep, press
,
then
Sync Setup. Then press Mode and Marker. Finally, press Marker
Freq. Then use the knob or numeric keypad to enter the desired
marker frequency.
Remote Interface Operation:
[SOURce[1|2]:]MARKer:FREQuency
{<frequency>|MINimum|MAXimum}

Table of Contents

Other manuals for Agilent Technologies 33521A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33521A and is the answer not in the manual?

Agilent Technologies 33521A Specifications

General IconGeneral
BrandAgilent Technologies
Model33521A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals