C
Copyright 2008 Agilent Technologies 
Agilent 4339B High Resistance Meter Operation Manual
Manual Change
Agilent Part No. N/A
Jun 2009
Change 1
Add TAR and change the Measurement Uncertainty value of Source Voltage Accuracy Test (Page 9-19) to
the following.
Source
Voltage
Setting
Test Limits Test Result
1
Measurement
Uncertainty
TAR
0 V ±0.10 V V ± 0.00021V
485
10 V ±0.12 V V ± 0.00027V
433
25 V ±0.14 V V ± 0.00054V
257
50 V ±0.18 V V ± 0.00096V
188
100 V ±0.26 V V ± 0.00185V
141
200 V ±0.42 V V ± 0.0037 V
115
201 V ±0.82 V V ± 0.0040 V
206
250 V ±0.90 V V ± 0.0053 V
170
500 V ±1.3 V V ± 0.0097 V
133
1000 V ±2.1 V V ± 0.0188 V
112
Change 2
Add TAR and change the Measurement Uncertainty value of Ammeter Offset Voltage Test (Page 9-19) to
the following.
Test Limits Test Result
1
Measurement
Uncertainty
TAR
±0.5 mV mV ± 0.0083 mV 60.27
Change 3
Add TAR and change the Measurement Uncertainty value of Ammeter Input Resistance Test (Page 9-19) to
the following.
Test Limits Test Result
1
Measurement
Uncertainty
TAR
±50 ohm ohm ± 4.37 ohm 11.5

Other manuals for Agilent Technologies 4339B