Getting
Started
4339B
T
o
Apply
the
T
est
V
oltage
The
following
commands
applies
the
test
voltage:
:OUTP
F
or
example
,
OUTPUT
717;":OUTP
ON"
:
OUTPUT
717;":OUTP
OFF"
W
arning
Sending
:OUTP
command
may cause
the
4339B
to
output
a
dangerous
voltage
levels up
to 1000
Vdc.
Do
NO
T
touch
the
UNKNO
WN
terminals
or
the
electrodes of
the accessory
,
when
the
V
Output
indicator
is
ON.
T
o
Set
the
Current
Limit
The
following
command
sets
the
current
limit
value:
:SOUR:CURR:LIM
F
or example
,
to
set
0.5
mA
as
the
current
limit
value
,
OUTPUT
717;":SOUR:CURR:LIM
0.5MA"
T
o
Select
Measurement
Time
Mode
The
following
command
selects
the
measurement
time
mode:
:SENS:CURR:APER
T
o
set
the
measurement
time
mode
to
SHORT:
OUTPUT
717;":SENS:CURR:APER
0.01"
T
o
set
the
measurement
time
mode
to
MEDIUM:
OUTPUT
717;":SENS:CURR:APER
0.03"
T
o
set the
measurement time
mode to
LONG:
OUTPUT
717:":SENS:CURR:APER
0.39"
T
o
P
erform
Calibration
The following command calibrates the 4339B:
:CAL?
F
or example
,
:
OUTPUT 717;":CAL?"
ENTER 717;A
IF A<>0 THEN
PRINT "Error:",A
PAUSE
END IF
:
4-6 Remote Operation