4339B
Getting
Started
T
o
Use
the
Comparator
Function
The
following
commands
control
the
comparator
function:
:CALC:LIM:CLE
:CALC:LIM:FAIL?
:CALC:LIM:LOW
:CALC:LIM:LOW:STAT
:CALC:LIM:UPP
:CALC:LIM:UPP:STAT
:CALC:LIM:STAT
F
or
example
,
to
set
the limit
values and
to enable
the comparator
function,
:
OUTPUT
717;":CALC:LIM:LOW
1E10"
OUTPUT
717;":CALC:LIM:UPP
1E11"
OUTPUT
717;":CALC:LIM:STAT
ON"
OUTPUT
717;":FETC?"
ENTER
717;S,D,C
:
T
o
Display
a
Deviation
Measurement
The following
commands display
deviation
results:
:DATA
:CALC:MATH:EXPR:NAME
:CALC:MATH:EXPR:CAT?
:CALC:MATH:STAT
F
or
example
,
to
set
the
reference
value
using
a
measurement
value
and
to
calculate
the
absolute
deviation
value
,
:
OUTPUT
717;":FETC?"
ENTER 717;s,d
OUTPUT
717;":DATA
REF,";d
OUTPUT
717;":CALC:MATH:EXPR:NAME
DEV"
OUTPUT
717;":CALC:MATH:STAT
ON"
:
To Set the Oset-Error
Canceling Function
The following command set the oset-error
canceling function ON or OFF:
:CAL:AUTO
For example
, to select the oset-error canceling function OFF
,
OUTPUT 717;":CAL:AUTO OFF"
ToWait Until Previously Sent Commands are Completed
The following commands make the 4339B wait until the previously sent commands are
completed:
*OPC
*OPC?
*WAI
Remote Operation 4-9