4339B
CALCulate
Subsystem
:CALCulate1:FORMat
f
REAL
j
SRESistivity
j
VRESistivity
g
Species
or
queries
the
post-processing
of
measurement
data,
which
is
measured using
the
<
sensor
function
>
specied
by
the
[SENSe]:FUNCtion
command. The
denition of
the SENSe
subsystem
parameters
are
as
follows:
<
sensor
function
>
of
[:SENSe]:FUNCtion
CALCulate1:FROMat
P
arameter
"CURRent:DC" REAL
current:
I
[A]
"RESistance[:DC]" REAL
resistance:
R
[]
SRESistivity
surface resistivity:
P
2
R
g
[]
VRESistivity
volume
resistivity:
A
t
2
R
[
2
cm
]
Where
,
P
arameter
Description Command
P
[m]
Eective
perimeter
CALCulate1:RESistivity:EPERimeter
g
[m]
Gap
between
main
electrode
and
guard
electrode (ring)
CALCulate:RESistivity:GLENgth
A
[m
2
] Eective
area
CALCulate1:RESistivity:EARea
t
[m]
Thickness
of sample
CALCulate1:RESistivity:STHickness
Query
response
is
REAL
,
SRES
,
or
VRES
.
:CALCulate1:LIMit:BEEP
er:CONDition
f
P
ASS
j
F
AIL
g
Denes
comparator
output
to
beeper
.
FAIL
a
beep
is
emitted
when
the
comparison
result
is
F
AIL
(High,
Low
or
No-contact).
PASS
a
beep
is
emitted
when
the
comparison
result
is
P
ASS
(In).
This
command
is
eective
when
SYSTem:BEEPer:STATe
and
CALCulate1:LIMit:BEEPer:STATe
commands
are
set
to
ON
.
Query
response
is
FAIL
or
PASS
.
:CALCulate1:LIMit:BEEP
er[:ST
A
T
e]
f
ON
j
OFF
j
1
j
0
g
Sets
or
queries
if comparator
output
to
beeper
is
enabled.
ON
or
1
Enables the comparator output
to beeper as dened by
:CALCulate1:LIMit:BEEPer:CONDition
command.
OFF
or
0
Disables comparator output to beeper
.
When you enable the beep function,
:SYSTem:BEEPer:STATe
command is automatically set to
ON
.
Query response is
1
or
0
.
:CALCulate1:LIMit:CLEar
Clears the data which is reported by
CALCulate1:LIMit:FAIL?
command. (no query)
GPIB Reference 5-11