4339B
FETCh?
Query
FETCh? Query
:FETCh?
Retrieves
a
measurement value
taken using
the
INITiate
subsystem commands
and places
the
measurement
data
into
the
4339B
's
output buer
. (Query
only)
Query
response
format
is
,
<
stat
>
,
<
data
>
,
<
comp
>
Where
,
<
stat
>
measurement
status
0
:
Normal
1:
Overload
2:
No-Contact
4
:
Over-Current (exceeding
current limit)
<
data
>
Measurement
data
<
comp
>
Comparison
result
(
no
output
when the
comparator
function
is
OFF)
1
:
In
2
:
High
4
:
Low
8
:
No-Contact
Data
format
is
specied
by
the
FORMat[:DATA]
command.