Measuring
Insulation
Resistance
of Capacitors
4339B
9.
Set the
measurement sequence
conguration.
Press
.
Select
Chrg
using
or and
press
.
Enter the
charge time
,
for
example 60
(s). Press
twice
.
10. Select
the
single
mode
as
the
measurement
sequence
mode
.
Press
.
Select
Single
using
or and
press
.
11.
Start
the
measurement
sequence
.
Press .
The
Seq
Running
annunciator(
9
)
is
turned ON,
and the
4339B applies
the test
voltage
across
the
capacitor
for
60
seconds
.
Note
If you
want to
abort
the
measurement
sequence
,
press
.
When
the sequence
measurement
is
nished,
the
annunciator
is
turned
OFF
and
the
measurement
result
is
displayed.
12.
T
o
exit
the
measurement
sequence
,
Press
.
Select
Off
using
or and
press
.
13. W
ait until
the
High
V
oltage
indicator
is
OFF
.
Then
remove
the
DUT
from
the
test
xture
.