4339B
Measuring
Resistivity
of Insulation
Materials
7. Turn
the load
knob to
apply pressure
by the
upper
and
lower
electrodes
on
the
DUT
,
and
then close
the top
cover.
Turning load
knob applies
load pressure
on
the
DUT
of
up
to
approximately
10
kg.
If
you
do not
need to
apply a
specic
pressure
,
DUT
must
at
least
rmly
held
by
the
upper and
lower electrodes
.
8. Set
the resistivity
cell
parameter
a.
Press
.
b. Enter
DUT thickness
.
Select
Thickenss
using
or and
press
.
Enter
the thickness
value,
for
example
0.13
(mm),
and
press
.
c. Select
the resistivity
cell size
.
Select
ResCell
using or and
press
.
Select
50
(mm)
using
or ,
and
press
twice
.
d.
Select
Exit
and
press
to
exit.
9.
Set
the
measurement
sequence
conguration.
Press
.
Select
Charge
and
press
.
Enter
the
charge
time
,
for
example
60
(s),
and press
twice.
10. Select the single mode
as the measurement sequence mode
.
Press
. Select
Single
using
or and press .
Application Measurement 6-7