Specications
4339B
Specications
These
specications
are
the
performance standards
or limits
against which
the instrument
is
tested.
When
shipped
from the
factory,
the 4339B
meets the
specications listed
in this
section.
The
specication
test
procedures are
covered in
Chapter 9.
Measurement
Parameters
R
(dc resistance),
I
(dc
current),
s
(surface
resistivity),
v
(volume
resistivity)
Measurement
Conditions
DC T
est
V
oltage
(Vs)
Range
0
to
1000
V
,
0.1
V
steps
for
Vs
200
V
or
1.0
V
steps
for Vs
>
200
V
A
ccuracy
6
(0.16%
+
100
mV)
for
Vs
200
V
6
(0.16%
+
500
mV)
for
Vs
>
200
V
Maximum Current
10
mA
for
Vs
100
V
5
mA
for
Vs
250
V
2
mA
for
Vs
500
V
1
mA
for
Vs
>
500
V
Current
Compliance
Setting
0.5
mA,
1
mA,
2
mA,
5
mA,
10
mA
Ammeter
Input
Resistance
1
k
6
5%
Input
Oset V
oltage
6
500
V
T
est
Cable
Length
2 meters maximum
Measurement Time Mode
Short, Medium, and Long
Ranging
Auto and Hold (manual)
Averaging
1to256
8-2 Specications