4339B
P
erformance
Tests
P
erformance
T
est
Record
Agilent
T
echnologies
4339B
High
Resistance
Meter
Serial
No.:
Date:
T
emperature:
T
ested by:
Humidity:
Source
V
oltage A
ccuracy T
est
Source V
oltage
Setting
Test
Limits T
est Result
1
Measurement
Uncertainty
0
V
6
0.10
V
V
6
0.000002
V
10
V
6
0.12
V
V
6
0.00018
V
25 V
6
0.14
V
V
6
0.00073 V
50
V
6
0.18
V
V
6
0.00073
V
100
V
6
0.26
V
V
6
0.0022
V
200
V
6
0.42
V
V
6
0.0043 V
201
V
6
0.82
V
V
6
0.0043
V
250
V
6
0.90
V
V
6
0.0028
V
500
V
6
1.3
V
V
6
0.0086 V
1000
V
6
2.1
V
V
6
0.011
V
1
T
est
Result
=
Measured
V
alue
0
Setting
V
alue
Ammeter
Oset
V
oltage
and
Input
Resistance
T
est
Ammeter
Oset
V
oltage
T
est
T
est
Limits
T
est
Result
Measurement
Uncertainty
6
0.5
mV
mV
6
0.0015
mV
Ammeter
Input Resistance
Test
Test Limits
Test Result
1
Measurement
Uncertainty
6
50
6
0.52
1
Test Result
= Typical V
alue of 1 k
0
Measured V
alue
Record-1 Maintenance 9-19