Agilent Technologies 4339B - Page 269

Agilent Technologies 4339B
287 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4339B
Procedure
T
able B-3.
Pull-up Resistor
Location
Socket
No
.
Signal
Name
Signal
Type
J11-1 /INDEX Control Signal
(5 to
15 V)
J11-2 /EOM
J11-3 /ALARM
J11-4 /NOT
READY
J11-5 /NO
CONT
A
CT
Comparison
Signal
(5
to
24
V)
J11-6 /LO
J11-7 /HI
J11-8 /IN
J11-9 Not
Used
J11-10
J11-11
6. Mount
the
pull-up
resistors
for
the
comparison
output
signals
.
(Refer
to
Figure
B-5
and
Table
B-3 for
the
location
of
the
pull-up
resistors
for
the
comparison
output
signals
.)
Use
the
following equation
to
determine
the
value
of
the
pull-up
resistors
(R).
R
[k]
'
Vp
[V]
/
3
where
,
Vp
is
the
pull-up
voltage
.
The
typical
pull-up
resistor
values
are:
Pull-up
V
oltage
Pull-up Resistor
Agilent
P
art
Number
5
V
0757-0278
(1.78 k)
12
V
0757-0279
(3.16
k)
24
V
0757-0441
(8.25
k)
7.
Mount
the
pull-up
resistors
for
the
control
output
signals
.
(Refer
to
Figure
B-5
and
T
able
B-3
for
the
location
of
the
pull-up
resistors
for
the
control
output
signals.)
Use the
following
equation
to
determine
the
value
of
the
pull-up
resistors
(R).
R
[k]
'
Vp
[V]
/
2.5
where, Vp is the pull-up voltage
.
The typical
pull-up resistor values are:
Pull-up
Voltage
Pull-up Resistor
Agilent P
art Number
5V 0757-0278 (1.78 k)
9V 0757-0279 (3.16 k)
12 V 0698-3154 (4.22 k)
15 V 0757-0438 (5.11 k)
8. Set SW1 according to the voltage value of EXT DCV2.
Handler Interface Installation B-9