4339B
GPIB
Errors
-211
Trigger ignored
A GET
,
*TRG
,or
triggering signal
was received
and recognized
by the
4339B
but
was
ignored
because of
4339B timing
considerations; for
example,
the 4339B
was
not
ready
to
respond.
-213
Init
ignored
A
request
for
a
measurement
initiation
was
ignored as
another measurement
was already
in
progress
.
-221
Settings
conict
A
legal
program
data
element
was
analyzed
syntax
but
could
not
be
executed
due
to the
current
device
state
(see
IEEE
488.2,
6.4.5.3
and
11.5.1.1.5).
-222
Data
out
of
range
A
legal
program
data
element
was
analyzed
syntax
but
could
not
be
executed
because
the
interpreted
value
was
outside
the
legal
range
as
dened
by
the
4339B
(see
IEEE
488.2,
11.5.1.1.5).
-223
T
oo much
data
A
legal
program
data
element
of
block,
expression,
or
string
type
was
received that
contained
more
data
than
the
4339B
could
handle
due
to
memory
or
related
device-specic
requirements
.
-230
Data
corrupt
or
stale
P
ossibly
invalid
data;
new
reading started
but
not
completed
since
last
access
.
-241
Hardware
missing
A
legal
program
command
or
query
could
not
be
executed
because
of
missing
4339B
hardware;
for
example
,
an
option
was
not
installed.
-310
System
error
Some
error
, termed
\system error"
by the
4339B
,
has
occurred.
-311
Memory
error
An
error was
detected
in
the
4339B
's
memory
.
-313
Calibration memory lost
The nonvolatile calibration data has been lost. When this error occurs during power-on test,
the 4339B stops operation and asserts the /ALARM signal on the handler interface
. When
this
error occurs during self-test, the 4339B continues operation and asserts the /ALARM signal.
Contact your nearest Agilent T
echnologies oce
.
-350
Queue
overow
A specic code entered into the queue in lieu of the code that caused the error. This code
indicates that there is no room in the queue and an error occurred but was not recorded.
Messages-7