EasyManua.ls Logo

Agilent Technologies 4339B - Page 97

Agilent Technologies 4339B
287 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4339B
Front
P
anel
1.
At the
Ready state
, pressing
starts
a measurement
sequence,
and the
test voltage
is
turned
ON.
2.
The 4339B
waits for
the dened
Charge Time
while
the
applied
voltage
is
charging
the
DUT
.
3.
After that,
the 4339B
repeats
measuring
the
DUT
as
many
times
as
is
dened
by
the
Memory
Size
with
a dened
Interval Time
between
measurements
.
The
measurement
data
is
displayed on
the
LCD
display
after
each
measurement,
and
all
measurement
data
is
stored
in
the data
buer
,
which
can
be
accessed
using
GPIB
commands
(refer
to
\D
A
T
A
Subsystem
"
in
Chapter 5).
4.
After all
measurements
are
completed,
the
4339B
turns
the
test
voltage
OFF
,
and
returns
to
the
Ready state
.
The
user
denable
parameters
Charge
Time
,
Interval
Time
,
and
Memory
Size
,
are
set
using
the
program
key
(
).
F
or
details
,
refer
to
\Program
Key
".
When
the
sequence
is
running,
the
Seq
Running
annunciator(
9
)
turns
ON.
Program
K
ey
The
program key
sets the
parameters for
a measurement
sequence
.
F
or
details
about
the
measurement
sequence
,
refer to
\Measurement Sequence
Mode
Key
".
Y
ou
can setup
the
measurement
sequence
parameters
in the
following menu:
Chrg
The
Charging
time
of
voltage
.
The
range
and
resolution
are
as
follows:
Range Resolution
0
s
Charge
Time
<
10
s
1
ms
10
s
Charge
Time
<
100
s
10
ms
100
s
Charge
Time
999
s
100
ms
The
default
setting
is
0
ms
.
Intvl
This
menu
is
for
setting
the
Interval
Time
and
the Memory
Size.
These
parameters
are
used
only
in
the
Continuous
mode
.
Interval Time
The
range
and
resolution
are
as
follows:
Range Resolution
10 ms
Interval Time
<
10 s
1ms
10 s
Interval Time
<
100 s
10 ms
100 s
Interval Time
999 s
100 ms
The default setting is 1 s
.
Memory Size
This parameter is the number of measurement points which will be stored.
Available memory sizes are 1 to 500. The default memory size is 500.