EasyManua.ls Logo

Agilent Technologies 4339B - Page 99

Agilent Technologies 4339B
287 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4339B
Front
P
anel
Parameter Description Available
Value
D
1
Main
electrode diameter
26
mm
1
50
mm
1
76
mm
1
D
2
Guard
ring diameter
38
mm
1
70
mm
1
88
mm
1
B
Eective
area coecient
0
B
1
(0.0001
resolution)
1
These parameter's
values
are
xed
and
cannot
be
changed.
Surface
resistivity (
s
)
and volume
resistivity
(
v
)
are calculated
using
the
following
equation:
s
=
(
D
2
+
D
1
)
(
D
2
0
D
1
)
2
R
s
[]
v
=
2
D
1
+
B
(
D
2
0
D
1
)
2
2
4
t
2
R
v
10
[
1
cm
]
Where
,
R
s
:
Surface
resistance
(Measured
value)
R
v
:
V
olume
resistance
(Measured
value)
t
:
DUT's
thickness
(Entered
as
Thickness
)
User
Denes
the
electrode's
parameters
when
using
your
own
test
xture
.
P
arameter
Description A
vailable
V
alue
Area
Eective
area
0
cm
2
AREA
9999.9
cm
2
(5-digit
number
,
default
is
19.635 cm)
P
erimeter
Eective
perimeter
0
cm
PERIMETER
999.99
cm
(5-digit
number
,
default
is
18.850
cm)
Gap
Gap 0.001
cm
GAP
99.99
cm
(4-digit number
,
default
is
1
cm)
Surface
resistivity (
s
)
and volume
resistivity
(
v
)
are calculated
using
the
following
equation:
s
=
PE
R
I
M
E
T
E
R
GAP
2
R
s
[]
v
=
ARE A
t
2
R
v
10
[
1
cm
]
Where,
R
s
: Surface resistance
R
v
: Volume resistance
t
: DUT's thickness (entered as
THICKNESS
)
Note
The eective area coecient
B
can be set only with front panel keys.Itis
impossible to enter the
B
value using GPIB commands.
Function Reference 3-9

Other manuals for Agilent Technologies 4339B