Selecting
an
appropriate
connection of
DUT
F
or
Impedance
Measurement
(Option
010)
Connecting
the Impedance
Test
Kit
In
the impedance
measurement,
the
43961A
Impedance
T
est
Kit
is
required
to
connect
your
DUT
to the
analyzer
.
See
Figure
5-9
.
1.
V
erify
the
4395A
is
turned
o.
2.
Connect
the
N-cable
to
the
RF
OUT
port
of
the
analyzer
.
3.
Connect
two
connectors
of
the
43961A
to
the
R
and
A
ports
of
the 43961A.
4.
Connect
the
other
connector
of
the
N-cable
to
the
RF
IN
port
of
the 43961A.
5.
Turn
on
the
4395A
.
Figure 5-9.
Connecting
the
Impedance
T
est
Kit
Note
The connection
between DUT
and the
analyzer is
accomplished when
atest
xture
is
mounted
on
the
impedance
test
kit.
Note
that
all
the
measurement
settings and calibrations should be completed before you mount the test xture
.
Follow the steps described in Chapter 6
and Chapter 7.
5-8 Preparations for Measurements