Selecting
the
A
ctive Channel
Selecting the
Analyzer Mode
1.
Press
4
Meas
5
.
2.
Press
N
N
N
N
N
N
N
NN
NN
NN
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
ANALYZER
TYPE
.
3.
Select one
of the
following
analyzer
modes:
Analyzer
Mode
Softkey
Network
Analyzer
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
NETWORK
ANALYZER
Spectrum
Analyzer
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
SPECTRUM
ANALYZER
Impedance
Analyzer
1
N
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
IMPEDANCE
ANALYZER
1
Option
010 only
.
Note
When
you
change
the
analyzer
mode
(operating
mode),
the
analyzer
is
set to
the
preset
state
.
So
,
when
you
want
to change
the analyzer
mode
,
you
must
select
it
before
you
set
up
the
other
settings
.
Note
In
this
manual,
the
following
abbreviations
are
used:
NA
mode
:
Network
analyzer
mode
S
A
mode
:
Spectrum
analyzer
mode
ZA
mode
:
Impedance
analyzer
mode
Selecting
the A
ctive Channel
In
the
A
CTIVE
CHANNEL block,
press
4
Chan
1
5
(channel
1)
or
4
Chan
2
5
(channel
2)
to
select the
active
channel.
The 4395A
has two
independent
channels
.
Y
ou
can
assign
a
dierent
set
of
measurement
conditions to
each channel
and switch
between
the
two
channels
with
just
a
single
keystroke
,
provided that
you use
both channels
for
the
same
analyzer
mode
.
By
switching
between
the
channels,
you can
eectively evaluate
the DUT's
characteristics under
dierent
measurement
conditions
.
Note
Be sure to select the active channel before you set measurement conditions
.
Once you have activated either channel, all the settings you make are
assigned to the active channel.
These two
channels can only be assigned dierent measurement
conditions within
the same analyzer mode
.Y
ou cannot use them across
two dierent analyzer
modes.F
or example, it is not possible to assign
network measurement conditions to
one channel while assigning spectrum
measurement conditions to the other channel.
6-2 Setting and Optimizing Measurement Conditions