Loading...
4395A
Option
010
Impedance Measurement
4395A Option
010 Impedance
Measurement
The
following
specications
are
applied when
the 43961A
Impedance T
est Kit
is connected
to
the
4395A
.
Measurement Functions
Measurement
parameters
Z,
Y
,
L,
C,
Q,
R, X,
G, B
,
Display
parameters
j
Z
j
,
z
,
R,
X,
j
Y
j
,
y
,
G,
B
,
j
0
j
,
,
0
x
,
0
y
,
Cp
,
Cs
,
Lp
,
Ls
,Rp
,Rs
,D
,
Q
Display
F
ormats
V
ertical
lin/log
scale
Complex
plane
P
olar/Smith/admittance
chart
Sweep P
arameters
Linear
frequency
sweep
Logarithmic
frequency
sweep
List
frequency
sweep
P
ower
sweep
(in
dBm
unit)
IF
Bandwidth
2,
10,
30,
100,
300,
1k,
3k,
10k,
30k
[Hz]
Calibration
OPEN/SHORT/LO
AD
3
term
calibration
Fixture
compensation
P
ort
extension
correction
Measurement
P
ort Type
APC-7
Output Characteristics
Frequency range
::::: :: :: ::
::::: :: :: :: ::: :: :::: ::: ::
::::::::::::::::::::::::
100
kHz to 500 MHz
Frequency resolution
::::: :: :: :: ::: :: :::: :
:::::::::::::::::::::::::
::: :: ::: :: :: :: ::::: :: :: :
1
mHz
Output impedance
:::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
50 nominal
Output Level
when the measurement port is terminated by 50
1
:::: :: :: :: ::: :: :::: ::: :: :
0
56 to +9 dBm
when the measurement port is open
:::::::::::::::::::::::::::::::
0.71 mVrms to 1.26 Vrms
1
Note: When the measurement port is terminated with 50 , the signal level at the
measurement port is 6 dB lower than the signal level at the RF OUT port.
11-14 Specications and Supplemental Characteristics