System
performance
with
Dierent T
est Sets
and Connector
Types
T
able 11-6.
Typical
System P
erformance
for
Devices
with
75
Type-N
Connectors
4395A
with 87511B
T
est
Set
(300
kHz
to
500
MHz)
Symbol Error
T
erms
Uncorrected
Typical Residual
after
A
ccuracy
Enhancement
1
,
2
Response
Only
Response
and
Isolation
One-P
ort
Full
two
port
D Directivity
0
33
dB
(2.2
2
10
0
2
)
0
33
dB
(2.2
2
10
0
2
)
0
44
dB
3
(6.3
2
10
0
3
)
0
44
dB
(6.3
2
10
0
3
)
0
44
dB
(6.3
2
10
0
3
)
M
s
Source
Match
0
20
dB
(0.10)
0
20
dB
(0.10)
0
20
dB
(0.10)
0
35
dB
(0.02)
0
35
dB
(0.02)
T
r
Reection
Tracking
6
3dB
(0.41)
6
1dB
(0.12)
6
0.9 dB
(0.11)
6
0.06 dB
(6.9
2
10
0
3
)
6
0.06 dB
(6.9
2
10
0
3
)
M
l
Load
Match
0
20
dB
(0.10)
0
20
dB
(0.10)
0
20
dB
(0.10)
{
0
42
dB
(7.9
2
10
0
3
)
T
t
Trans
.
Tracking
6
3
dB
(0.41)
6
0.1
dB
(0.01)
6
0.1
dB
(0.01)
{
6
0.05
dB
(6.9
2
10
0
3
)
C Cross T
alk
0
100
dB
(1.0
2
10
0
5
)
0
100
dB
(1.0
2
10
0
5
)
0
104
dB
(6.3
2
10
0
6
)
{
0
104
dB
(6.3
2
10
0
6
)
R
r1
P
ort1
Re.
Connector
Repeatability
0
65
dB
(5.6
2
10
0
4
)
R
t1
P
ort1 Trans
. Connector
Repeatability
0
65 dB
(5.6
2
10
0
4
)
R
r2
Port2
Re. Connector
Repeatability
0
65
dB (5.6
2
10
0
4
)
R
t2
P
ort2
Trans
. Connector
Repeatability
0
65
dB
(5.6
2
10
0
4
)
N
l
Low-Level
Noise
2
0
104
dB
from
full
scale
(6.3
2
10
0
6
)
N
h
High
Level
Noise
2
,
4
Magnitude:0.001
dB
(1.2
2
10
0
4
)
A
m
,A
p
Dynamic
A
ccuracy
Error
See
\Dynamic
A
ccuracy"
in
Chapter
11.
U
m
,U
p
Multiplexer
Switching
Uncertainty
Magnitude:
6
0.002
dB
(2.3
2
10
0
4
)
Phase:0.015
degrees
S
t1
P
ort1
Cable
Trans
.
Phase
Stability
5
0.05
2
f
(GHz)
degrees
S
r1
P
ort1
Cable
Re.
Stability
5
0
70
dB
(3.2
2
10
0
4
)
S
t2
P
ort2
Cable
Trans
.
Phase Stability
5
0.05
2
f
(GHz)
degrees
S
r2
P
ort2
Cable
Re.
Stability
5
0
70
dB
(3.2
2
10
0
4
)
T
td
Trans
.
Tracking
Drift
Magnitude:
0.01
dB/
C
(1.2
2
10
0
3
/
C)
Phase:
6
6
[0.1+0.15
2
f(GHz)]degrees/
C
T
rd
Re. Tracking Drift Magnitude: 0.01 dB/
C (1.2
2
10
0
3
/
C) Phase:
6
6
[0.1+0.15
2
f(GHz)]degrees/
C
T
sw
Switch Tracking
6
0.03 dB (3.5
2
10
0
3
)
M
sw
Switch P
ort Match
0
70 dB (3.2
2
10
0
4
)
1
Accuracy enhancement procedures are performed using 85036B 75 type-N calibration kit. Enviromental temperature
is 23
C
6
3
C at calibration:
6
1
C from calibration temperature must be maintained
for valied measurement calibration.
2
With IF bandwidth of 10 Hz.
3
With impedance matched load.
4
High-level noise is the RMS of a continuous measurement of a short circuit or thru.
5
Arrived at by bending 11857D cables out perpendicular to front panel and reconnecting. Stability is much better with
less exing.
6
Arrived at using 11857D cables and full 2-port calibration. Drift is much better without calbes and with 1-port
calibration. For this case, drift typically is [0.1 + 0.05
2
f (GHz)]
2
1
C, degrees.
Specications and Supplemental Characteristics 11-49