Calibration
for
Network
Measurement
Frequency
Response
(Tracking)
This
is
the
vector
sum
of
all
test
setup
variations
in which
magnitude and
phase change
as a
function
of
frequency
.
This
includes
variations contributed
by signal
separation devices
, test
cables
,
and
adapters
,
and
variations between
the reference
and test
signal paths
.This
error is
a
factor
in
both
transmission
and
reection
measurements.
F
or
further
explanation
of
systematic
error
terms
and
the
way they
are combined
and
represented
graphically
in
error
models
,
see the
later section,
titled A
ccuracy Enhancement
Fundamentals
-
Characterizing
Systematic
Errors
.
Compensation
for
Measurement
Errors
There
are
twelve
dierent
error
terms
for
a
two-port
measurement
that
can
be
corrected
by
accuracy
enhancement
in
the
analyzer
.
These
are
directivity
,
source
match,
load
match,
isolation,
reection
tracking,
and
transmission
tracking,
each
in
both
the
forward
and
reverse
direction.
The
analyzer
has
several
dierent
measurement
calibration
routines
to
characterize
one
or
more
of
the
systematic
error
terms
and
remove
their
eects
from
the
measured
data.
The
procedures
range
from
a
simple
frequency
response
calibration
to
a
full
two-port
calibration
that
eectively
removes
all
twelve
error
terms
.
The
Response
Calibration
eectively
reduces the
frequency
response
errors
of
the
test
setup
for reection
or transmission
measurements.
This calibration
procedure
may
be
adequate
for
measurement
of
well-matched
low-loss
devices
.
This
is
the
simplest
error
correction
to
perform,
and
should
be
used
when
extreme
measurement
accuracy
is
not
required.
The
Response
and
Isolation
Calibration
eectively
removes
frequency
response
and
crosstalk
errors
in
transmission
measurements
,
or
frequency
response
and
directivity
errors
in
reection
measurements
.
This
procedure
may
be
adequate
for
measurement
of
well-matched
high-loss
devices
.
The
S
11
and
S
22
One-P
ort
Calibration
procedures
provide
directivity
,
source
match,
and
frequency
response
vector
error
correction
for
reection
measurements
.
These
procedures
provide
high
accuracy
reection
measurements
of
one-port
devices
or
properly
terminated
two-port
devices
.
The
Full
Two-P
ort
Calibration
provides directivity
, source
match, load
match, isolation,
and
frequency
response
vector
error
correction,
in
both forward
and reverse
directions,
for
transmission
and
reection
measurements
of
two-port
devices
. This
calibration provides
the
best
magnitude
and
phase
measurement
accuracy
for both
transmission and
reection
measurements
of
two-port
devices
,
and
requires
an S-parameter
test set.
The
One-P
ath
Two-P
ort
Calibration
provides
directivity
,
source match,
load
match,
isolation,
and frequency response vector error correction in one direction. It is used for high accuracy
transmission and reection measurements using a transmission/reection test kit, such as the
87512A, B . (The DUT must be manually reversed between sweeps to accomplish measurements
in both the forward
and reverse directions
.)
All the calibration procedures described
above are accessed from the
4
CAL
5
key.
A-44 Basic Measurement Theory