Calibration
for
Network
Measurement
Figure
A
-30.
Reection
Tracking
E
RF
How
calibration
standards
are used
to quantify
these error
terms
.
It
can
be
shown
that
these
three
errors
are mathematically
related to
the actual
data,
S
11A
,
and
measured
data,
S
11M
,
by
the
following
equation:
S
11
M
=
E
D
F
+
S
11
A
(
E
R
F
)
1
0
E
S
F
S
11
A
If
the
value
of these
three
\E"
errors and
the measured
test
device
response
were
known
for
each
frequency,
the
above
equation
could
be
solved
for
S
11A
to
obtain the
actual
test
device
response
.
Because
each
of
these
errors
changes
with
frequency
,
their
values
must
be
known
at
each
test
frequency
.
These
values
are
found
by
measuring
the
system
at
the
measurement
plane
using three
independent
standards
whose
S
11A
is
known at
all
frequencies
.
The
rst
standard
applied
is
a
\perfect
load"
that
makes
S
11A
=
0
and
essentially
measures
directivity
(
Figure
A
-31
).
\P
erfect
load"
implies
a
reection-free
termination
at
the
measurement
plane
.
All
incident
energy
is
absorbed.
With
S
11A
=
0
the
equation
can
be
solved
for
E
DF
,
the
directivity
term.
In
practice
,
of
course
,
the
\perfect
load"
is
dicult
to
achieve
,
although
very
good
broadband
LO
ADs
are
available
in
the
4296A
compatible
calibration
kits
.
Figure A
-31. \P
erfect Load" T
ermination
Because the measured value for directivity is
the vector sum of the actual directivity plus
the actual reection coecient of the \perfect load," any reection from the termination
represents an error. System eective directivity becomes the actual reection coecient of the
\perfect load" (Figure A-32). In general, any termination having a return loss value greater
than the uncorrected system directivity reduces reection measurement uncertainty.
Basic Measurement Theory A-51