Agilent 4395A Network/Spectrum/Impedance Analyzer Operation Manual
Manual Change
Agilent Part No. N/A
January 2007
Change 1
Change the equations for the noise level measurement (page 8-35) to the following.
Converting to a Different Equivalent Noise Bandwidth
1. Calculate the conversion factor using the following equations with displayed units:
Unit Use
dBm/Hz K=10logBW
dBV/
√
Hz
、
dBμV/
√
Hz K=10logBW
W/Hz K=BW
V/
√
Hz K=
√
BW
Where, BW is the target equivalent noise bandwidth.
Change 2
Change options 1CN, 1CM and 1CP in the package contents of the 4395A (page 2-3) to the following.
Description Agilent Part Number Quantity
Option 1CN Handle Kit
Handle Kit
5063-9229 1
Option 1CM Rack Mount Kit
Rack Mount Kit
5063-9216 1
Option 1CP Rack Mount & Handle Kit
Rack Mount & Handle Kit
5063-9223 1