Sweep
Block
4
T
rigger
5
Function Range Resolution Preset V
alue
Power
ON default
Trigger
type
Continuos
, Number
of groups
,
Single,
Hold,
Continuous Continuous
NA:
Trigger
Source
Free
run,
External,
Manual, GPIB
Free
run
Free
run
NA:
Trigger
event
On
point,
On
sweep
On
sweep
On
sweep
Trigger polarity Positive
, Negative
Positive Positive
Gate
type
Level,
Edge
level Level
Gate
delay
0.8
s
to
3.2
s
0.8
s
0.8
sec
Gate length 6
sto
3.2 s
100
s
100
sec
SA:
Trigger
Source
Free
run,
External,
Manual,
GPIB
,
Level
Gate
1
,
Edge
Gate
1
Free
run
Free
run
1
With option
1D6.
4
Center
5
Function Range Resolution Preset
V
alue
P
ower
ON
default
NA:
Center
Frequency
0
Hz
to
510
MHz
1
mHz
250
MHz
250
MHz
NA:
Step
Size
mode
On(1-2-5
step),
O(linear)
On On
SA:
Center
Frequency
10
Hz
to
510
MHz
1
mHz
250
MHz
250
MHz
SA:
Step
Size
mode
On(1-2-5
step),
O(linear)
O O
Center
step
size
1
mHz
to
510
MHz
1
mHz
10
MHz
10
MHz
ZA:
Step
Size
mode
On(1-2-5
step),
O(linear)
On On
4
Span
5
Function Range Resolution Preset V
alue
Power ON default
NA: Span Frequency 0 Hz to 509.99999 MHz 1mHz 499.99999 MHz 499.99999 MHz
NA: Span power 0to20
dB
20 dB 20 dB
SA: Span Frequency 0 Hz to 510 MHz 1mHz 500 MHz 500 MHz
ZA: Span
Frequency
0Hz
to 509.99999 MHz
1mHz 499.99999 MHz 499.99999 MHz
C-16 Input Range and Default Settings