Agilent Technologies 4395A - Page 543

Agilent Technologies 4395A
575 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error
Messages
in
Alphabetical Order
0
108
P
arameter not
allowed
More parameters
were received
than expected
for the
header.
For
example,
the
*SRE
command
only accepts
one parameter
,so
receiving
*SRE
4,16
is
not
allowed.
48
PHASE
LOCK LOOP
UNLOCKED
EXT
REF
Input
of
10
MHz
is
not proper
,or
the instrument
is needed
to
adjust
or
repair
.
Check
the
external
reference
signal
rst.
Contact
your nearest
Agilent T
echnologies oce
for
adjustment
or
repair
.
193
PO
WER
ON TEST
FAILED
P
ower
on
test
failed.
Contact
your
nearest
Agilent
T
echnologies oce
.
26
PRINTER:
not
on,
not
connected,
out
of
paper
The
printer
does
not
respond
to
control.
Check
the
supply
to
the
printer
,
online
status
,
sheets
,
and
so
on.
0
284
Program currently
running
Certain
operations dealing
with
programs
may
be
illegal
while
the
program
is
running.
F
or
example
,
deleting
a
running
program
might
not
be
possible
.
0
280
Program
error
Indicates
that
a
downloaded
program-related
execution error
occurred.
This
error
message
is
used
when
the
analyzer
cannot detect
the
more
specic
errors
described
for
errors
0
281
through
0
289.
0
112
Program
mnemonic
too
long
The
header
contains
more
than
twelve
characters
(see
IEEE
488.2,
7.6.1.4.1).
0
286
Program
runtime
error
A
program
runtime error
of the
HP Instrument
B
ASIC
has
occurred.
T
o
get
a
more
specic
error
information,
use the
ERRM$
or
ERRN
command
of
the
HP
Instrument
B
ASIC.
0
285
Program
syntax
error
Indicates
that a
syntax
error
appears
in
a
downloaded
program.
The
syntax
used
when parsing
the downloaded program is device-specic
.
Q
0
400
Query errors
This is the generic query error that the analyzer cannot detect more specic errors
. This code
indicates only that a query error as dened in IEEE 488.2, 11.5.1.1.7 and 6.3 has occurred.
0
410
Query INTERRUPTED
A condition causing an interrupted query error occurred (see IEEE 488.2, 6.3.2.3). For
example, a query followed by
DAB
or
GET
before a response was completely sent.
Messages-11