Replacing
Fuse
Replacing Fuse
Fuse
Selection
Select
proper
fuse
according
to
the
T
able
2-2
.
T
able
2-2.
Fuse
Selection
Fuse
Rating/Type
Fuse
P
art
Number
5A 250V
ac
UL/CSA
type
2110-0030
Time Delay
F
or
ordering the
fuse,contact
your nearest
Agilent
T
echnologies
Sales
and
Service
Oce
.
Procedure
Lever
a
small
minus
screwdriver
to
dismount
the fuse
holder
above
the
A
C
line receptacle
on the
rear panel.
T
o
check
or
replace
the
fuse
,pull
the fuse
holder
and remove
the
fuse
.
T
o
reinstall
the fuse
, insert
afuse
with the
proper
rating
into
the
fuse
holder
.
Caution
Use the proper fuse for
the line voltage selected. Use only fuses with the
required current rating and of the specied type as replacements
. DO NOT use
a mended fuse or short-circuit the fuse-holder in order to by-pass a blown fuse
.
Find out what caused the fuse to blow!