5-40
Making Measurements
Setting measurement thresholds
4Press the Lower softkey, then turn the Entry knob to set the lower
measurement threshold value.
Increasing the lower value beyond the set middle value will automatically
increase the middle value to be more than the lower value. The default lower
threshold is 10% or 800 mV.
If threshold Typ e is set to %, the lower threshold value can be set from 5% to 93%.
5Press the Middle softkey, then turn the Entry knob to set the middle
measurement threshold value.
The middle value is bounded by the values set for lower and upper thresholds.
The default middle threshold is 50% or 1.20 V.
If threshold Typ e is set to %, the middle threshold value can be set from 6%
to 94%.
6Press the Upper softkey, then turn the Entry knob to set the upper
measurement threshold value.
Decreasing the upper value below the set middle value will automatically
decrease the middle value to be less than the upper value. The default upper
threshold is 90% or 1.50 V.
If threshold Ty pe is set to %, the upper threshold value can be set from 7%
to 95%.

Table of Contents

Other manuals for Agilent Technologies 54622D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54622D and is the answer not in the manual?

Agilent Technologies 54622D Specifications

General IconGeneral
BrandAgilent Technologies
Model54622D
CategoryTest Equipment
LanguageEnglish

Related product manuals