10 Verification
Table 2-2. Test Equipment Required for Verification
Type Required Characteristics Recommended Model
Current Monitor Resistor
100 A (0.01 ) ±0.04% for Agilent
Guildline 9230/100
6541A, 6551A, 6552A, 6641A,
6651A, & 6652A models.
15 A (0.1 ) ±0.04% for
Agilent
Guildline 9230/15
6542A, 6543A, 6544A, 6545A,
6553A, 6554A, 6555A, 6642A,
6643A, 6644A, 6645A, 6653A,
6654A, 6655A models.
DC Power Supply
5 V @ 10 A Agilent 66 Agilent 42A, 6653A
Digital Voltmeter
Resolution: 10 nV @ 1 V Agilent 3458A
Readout: 8 1/2 digits
Accuracy: 20 ppm
Electronic Load
Voltage and current range must
exceed range of supply under test.
Power range: 250 W minimum
Agilent
N3300A mainframe with
A
gilent
N3306A (60 V) plug-in
m
odule or Agilent
N3305A (150 V)
plug-in module.
GPIB Controller
1
Full GPIB capabilities Agilent 82350B
Load Resistor
0.1 ±5%, 300 W for Agilent
6541A, 6641A, 6551A, 6651A,
6552A, 6652A models.
Ohmite C300KRlO
l.0 ±5%, 300 W for Agilent
6542A, 6543A, 6544A, 6545A,
6553A, 6554A, 6555A, 6642A,
6643A, 6644A, 6645A, 6653A,
6654A, 6655A models.
Ohmite C300KlRO
Oscilloscope
Sensitivity: 1 mV Agilent
Infiniium
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip
RMS Voltmeter
True RMS Bandwidth: 20 MHz Agilent 3400B
Sensitivity: 100 µV
Variable-Voltage Transformer
Adjustable from -13% to +6% of
input range. Power: 1 kVA
minimum.
1
For 664xA and 665xA models only.