8 – Language Dictionary
110
Command Syntax
OUTPut:COMPensation:MODE <setting>
Parameters
LLOCAL | LREMOTE | HLOCAL | HREMOTE
*RST Value
HREMOTE
Examples
OUTP:COMP:MODE HREMOTE
Query Syntax
OUTPput:COMPensation:MODE?
Returned Parameters
<CRD>
Backward Compatibility
(Agilent 66311B/D, 66309B/D)
OUTP:TYPE[:CAPacitance] (LOW mode corresponds to
LLocal; HIGH mode corresponds to HRemote)
OUTPut:DFI
This command enables or disables the discrete fault indicator (DFI) output from the dc source.
Command Syntax
OUTPut:DFI[:STATe] <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
OUTP:DFI 1 OUTP:DFI ON
Query Syntax
OUTPut:DFI[:STATe]?
Returned Parameters
0 | 1
Related Commands
OUTP:DFI:SOUR
OUTPut:DFI:SOURce
This command selects the source for discrete fault indicator (DFI) events. The choices are:
QUEStionable
selects the Questionable event summary bit (bit 3 of the Status Byte Register)
OPERation
selects the Operation Event summary bit (bit 7 of the Status Byte Register)
ESB
selects the Standard Event summary bit (bit 5 of the Status Byte Register)
RQS
selects the Request Service bit (bit 6 of the Status Byte Register)
OFF
selects no DFI source
Command Syntax
OUTP:DFI:SOUR <source>
Parameters
QUES | OPER | ESB | RQS | OFF
*RST Value
OFF
Examples
OUTP:DFI:SOUR OPER
Query Syntax
OUTPut:DFI:SOUR?
Returned Parameters
<CRD>
Related Commands
OUTP:DFI
OUTPut:PON:STATe
This command selects the power-on state of the dc source. This information is saved in non-volatile
memory. The following states can be selected:
RST
Sets the power-on state to *RST. Refer to *RST for more information.
RCL0
Sets the power-on state to *RCL 0. Refer to *RCL for more information.
Command Syntax
OUTPut:PON:STATe <state>
Parameters
RST | RCL0
Examples
OUTP:PON:STAT RST
Query Syntax
OUTPut:PON:STATe?
Returned Parameters
<CRD>
Related Commands
*RST *RCL

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 66319D and is the answer not in the manual?

Agilent Technologies 66319D Specifications

General IconGeneral
BrandAgilent Technologies
Model66319D
CategoryPower Supply
LanguageEnglish

Related product manuals