EasyManua.ls Logo

Agilent Technologies 8163A - Page 150

Agilent Technologies 8163A
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurement Operations & Settings Signal Conditioning
150 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:INPut[n][:CHANnel[m]]:OFFSet:POWermeter
syntax: :INPut[n][:CHANnel[m]]:OFFSet:POWermeter<wsp><slot>,<channel>
description: Sets the offset factor (α
Offset
) to the difference between a power value measured by another
powermeter (hosted by the same mainframe) (P
ext
) and the power value measured by the
attenuator module’s monitor diode (P
att
).
α
Offset
(dB) = P
att
(dBm) - P
ext
(dBm)
parameters: <slot>
<channel>
Slot number of the external powermeter.
Channel number of the external powermeter.
response: none
example: INP1:OFFS:POW 4,2
affects: Attenuator modules with power control.
command:
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed
syntax: :INPut[n][:CHANnel[m]]:ATTenuation:SPEed<wsp><value> | MIN | MAX | DEF
description: Sets the filter transition speed; the speed at which the module moves from one attenuation to
another (in dBs).
parameters: <value>
MIN | MAX | DEF
The filter transition speed in dB/s.
Sets the filter transition speed to the module limits, or the module de-
fault.
response: none
example: INP1:ATT:SPE 2
affects: All attenuator modules.
command:
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed?
syntax: :INPut[n][:CHANnel[m]]:ATTenuation:SPEed?<wsp> MIN | MAX | DEF
description: Without the optional parameter, queries the transition speed of the filter.
parameters: MIN | MAX | DEF Queries the transition speed limits, or the module default.
response: 4 byte Intel floating point; transition speed in dB/s.
example: INP1:ATT:SPE? 2<END>
affects: All attenuator modules.