EasyManua.ls Logo

Agilent Technologies 8163A - Page 83

Agilent Technologies 8163A
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Root Layer Command Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 83
command:
:SLOT[n]:HEAD[m]:OPTions?
syntax: :SLOT[n]:HEAD[m]:OPTions?
description: Returns information about an optical head’s options.
parameters: none
response: A string.
example: slot1:head:opt? NO CONNECTOR OPTION, NO INSTRUMENT
OPTIONS<END>
affects: Optical heads
command:
:SLOT[n]:HEAD[m]:TST?
syntax: :SLOT[n]:HEAD[m]:TST?
description: Returns the latest selftest results for an optical head.
NOTE
This command does not perform a selftest. Use selfTeST command, “*TST?” on
page 62, to perform a selftest.
parameters: none
response: Returns an error code and a short description of the error.
example: slot:head:tst? +0,"self test OK"<END>
affects: Optical heads
command:
:SLOT[n]:HEAD[m]:WAVelength:RESPonse?
syntax: :SLOT[n]:HEAD[m]:WAVelength:RESPonse?
description: Returns the wavelength response from a wavelength calibrated module in binary format.
response: Wavelength Response table as a binary block.
response
format:
One 8 byte long wavelength calibration value pair consisting of a 4 byte long float for wave-
length and a 4 byte long float for the scalar calibration factor.
For more information on binary block formats see “Data Types” on page 26
example: slot1:head1:wav:resp? #536570........
affects: Attenuator with power control, all powermeters, return loss modules

Table of Contents