EasyManua.ls Logo

Agilent Technologies 8163A - Page 84

Agilent Technologies 8163A
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurement Operations & Settings Root Layer Command
84 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:CSV?
syntax: :SLOT[n]:HEAD[m]:WAVelength:RESPonse:CSV?
description: Returns the wavelength response from the attenuator module in CSV format.
response: Wavelength Response table as a string
response
format:
The string is a comma separated value (CSV) list and can be written to a file and be pro-
cessed with a spreadsheet program.
List format:
λ1, c1\n
λ2, c2\n
.......
λn, cn\n
"," separates wavelength and response factor
"\n" = ASCII code 10 separate value pairs
example: slot1:head1:wav:resp:csv? 1200e-6,2.019\n 1210e-6,
1.956\n...
affects: Attenuator with power control, all powermeters, return loss modules
command:
:SLOT[n]:HEAD[m]:WAVelength:RESPonse:SIZE?
syntax: :SLOT[n]:HEAD[m]:WAVelength:RESPonse:SIZE?
description: Returns the number of elements in the wavelength response table.
response Number of elements in the wavelength table as an integer value
example: slot2:head1:wav:resp:size? 50<END>
affects: Attenuator with power control, all powermeters, return loss modules
command:
:SPECial:REBoot
syntax: :SPECial:REBoot
description: Reboots the mainframe and all modules.
parameters: none
response: none
example: spec:reb