20.
FM Input
Impedance
(8360
B-Series
Only)
Description
and
Procedure
A
soldering
iron
is
required
for
this
pro
cedure.
This
pro
cedure
sets
the
FM
input
imp
edance
to
either
50
î™’
or 600
î™’. The
FM input
imp
edance
on
A11
must
be
set to
600 î™’
for
prop
er
op
eration
of
the
in
ternal
mo
dulation
generator
(Option 002).
Instruments
without
Option
002
should
p
erform
step
A,
\A11
Jump
er".
Instrumen
ts
with
Option
002
should
set
the
A11
jump
er
to
the
600
î™’
p
osition
using
step
A,
\A11
Jump
er"
and
should
p
erform
step
B,
\A8
Jump
er".
Agilent 8360 Adjustments
FM Input Impedance
3-89

Table of Contents

Other manuals for Agilent Technologies 8360 B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8360 B Series and is the answer not in the manual?

Agilent Technologies 8360 B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8360 B Series
CategoryPortable Generator
LanguageEnglish

Related product manuals