Page #111 background imageLoading...
Page #111 background image
Table
2-8. P
ow
er Accuracy
Frequencies
F
requency
(GHz)
Measured
Dierence
at
10
dBm
Measured
Dierence
at
>
10
dBm
0.1
1
1.0
1
6.0
10.0
18.0
23.0
1
28.0
1
36.0
1
45.0
1
1
Disregard
if
b
ey
ond
y
our
syn
thesizer's
capability
.
8.
On
the
p
o
w
er
meter,
set
the
p
o
w
er sensor
calibration factor
for the
frequency
to
b
e
measured.
9.
Note
the
dierence
bet
ween
the
p
o
w
er
meter
reading
and
the
p
o
w
er
v
alue
set
on
the
syn
thesizer.
W
rite
this
v
alue
do
wn
on
a
separate piece
of pap
er.
Note
F
or
p
ow
er
lev
els
of
10
dBm
and
ab
o
v
e,
measure
the
p
o
w
er
at
10
dBm,
then
set
the
atten
uator
to
the
10
dB
p
osition
(use
a
10 dB
atten
uator
for
syn
thesizers
without
a
step
atten
uator.)
Record
the
dierence
b
et
w
een
the
measuremen
ts:
(appro
ximately 10
dB).
Then,
with
the
atten
uator
still
set
to
10
dB,
complete
the
remaining
measuremen
ts.
Eac
hof
these measuremen
ts
should
b
e
oset
b
y
the
dierence
in
dB
that
y
ou
measured.
10. On
the
syn
thesizer,
use
the
up
4
*
5
k
ey
to
incremen
t
the
p
o
w
er
lev
el
1
dB.
11.
Rep
eat
steps 8
through 10,
to the
maxim
um
sp
ecied
p
o
w
er
lev
el
of
y
our
syn
thesizer.
Record
the
w
orst
case
measured
dierence
(greatest
duration)
for
this frequency
in
T
able
2-8
for
p
o
w
er
lev
els
10
dBm
and,
if
y
our
synthesizer
is capable,
for p
ow
er
lev
els
>
10
dBm.
12.
On
the
synthesizer,
set
the
p
o
w
er
lev
el
to
0
10
dBm.
13. Repeat steps 8 through 11 for the remaining frequencies in T
able 2-8
not to exc
eed
the
maximum specied frequency of y
our syn
thesizer.
Rememb er to delete the 10 dB of
attenuation for po
wer lev
els b elo
w0dBm.
14. Record the w
orst case measured v
alues for eac
h frequency range and
po
wer range on the
test record.
Related Adjustments
Mo dulator Oset and Gain
ALC Power Level Accuracy
ALC Power Flatness
2-24 Performance Tests
Power Accuracy
Agilent 8360

Table of Contents

Other manuals for Agilent Technologies 83650B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83650B and is the answer not in the manual?

Agilent Technologies 83650B Specifications

General IconGeneral
BrandAgilent Technologies
Model83650B
CategoryPortable Generator
LanguageEnglish

Related product manuals