17.
AM Bandwidth
(F
or
B-Series
Only)
Description
and
Procedure
This
test
v
eries
that
the
p
eak
RF
output
p
o
w
er
does
not v
ary more
than a
sp
ecied
amoun
t
o
v
er
the
sp
ecied amplitude
mo dulation
rate frequency
range. The
sp
ectrum
analyzer
with
the trac
king generator
op
erates
as
a
net
w
ork
analyzer
to
measure
the
atness.
Flatness
errors
asso
ciated
with
the
trac
king
generator
and
sp
ectrum
analyzer
are calibrated
out of
the
measuremen
t.
1. Connect
the
equipmen
t
as
sho
wn
in
Figure
2-26
with
the
trac
king
generator
output
connected
to
the
50
ohm
input
of
the
sp
ectrum
analyzer
and
to the
DVM.
Press
4
PRESET
5
and
let
the
equipment
warm
up for
at
least
one
hour.
Figure
2-26. AM
Bandwidth
T
est
Setup
Equipment
Spectrum Analyzer
With T
racking Generator
HP/Agilent 3585A/B
Crystal Detector
HP/Agilent 33330E (83640B/50B)
Crystal Detector
HP/Agilent 33330D (83620B/22B/23B/24B/30B)
Digital Voltmeter HP/Agilent 3456A
2. On the spectrum analyzer, set:
Start Frequency: 5kHz
Stop Frequency: 200 kHz
Sweep Manual: 5kHz
2-84 Performance Tests
AM Bandwidth
Agilent 8360

Table of Contents

Other manuals for Agilent Technologies 83650B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83650B and is the answer not in the manual?

Agilent Technologies 83650B Specifications

General IconGeneral
BrandAgilent Technologies
Model83650B
CategoryPortable Generator
LanguageEnglish

Related product manuals