46
N1010A FlexDCA and 86100A/B/C/D
Firmware Release Notes
A.01.22 Release Notes, September 6, 2000 (Differences from Release
A.01.21)
NOTE This firmware release is not supported.
New Feature List
•None
Defects Fixed
• The eye crossing algorithm is now more tolerant of heavily filtered eye waveforms.
The Start- Mask- Test will align to larger variety of eye waveforms. Eye
Measurements, which rely on an accurate eye crossing point, will perform better
on these filtered waveforms.
A.01.21 Release Notes, August 9, 2000 (Differences from Release
A.01.20)
NOTE This firmware release is not supported.
New Feature List
•None
Defects Fixed
•None
Factory- only Enhancement
• New routine added to increase the efficiency of the factory horizontal calibration
process. This improvement does not change the functionality of the product.
A.01.20 Release Notes, July 11, 2000 (Differences from Release
A.01.13)
NOTE This firmware release is not supported.
New Feature List
• External Scaling for vertical channel and trigger
• LAN capability for printing and file sharing
Defects Fixed
• With external keyboard connected, pressing ALT F4 allowed access to Windows
which is bad. This "feature" is now unavailable.
• Vertical cal temperature window is changed from >1 deg C to >5 deg C.
• Changed variable persistence default time to 300 ms.
• If the touchscreen is disabled and a message, such as "Autoscale Failed" is
displayed, the instrument will appear to be locked up because there is no way to