N1010A FlexDCA and 86100A/B/C/D
Firmware Release Notes
A.02.01)
New Feature List
New Return to Zero measurements
Added math function MIN/MAX
Added JPEG and TIFF File Support
Added scope measurements Vavg, Tmin, Tmax, Tedge, Delta Time
Added Histogram Measurement Peak
Ability to Show Histograms Window Boundary at all times when Histograms
feature is turned on
Added remote command to save screen image to hard or networked drive
Added remote command to save file in any supported format
Defects Fixed:
DCA crashes after default setup while in TDR mode
Remote command :MARKER:Y1POSITION? now correctly returns answer in volts
or ohms
Resolved measurement resolution issue, introduced in version A.02.01. In Eye
Mode on electrical channels the vertical scale resolution was fixed at 10 mV.
Fixed text on Limit Setup button doesn't reset
Filter rate values displayed correctly now
Instrument Hangs when trying to move Y- Axis Markers
Waveform colors no longer changed when invert waveform background color is
selected
Mask save screen command no longer overwrites previous data
The remote command :MTESt:SCALe:SOURce query now returns SCPI standard
character text, e.g. "CHAN1"
Fixed a release assertion problem when mapping a Sun Unix drive
Disk error message addressed because file overwriting itself
Long Filenames entered from File\Save menu causes problem
Waveform memory2 tab highlighted but information is about Waveform memory1
Resolved a problem with mask load & reset after print causes illegal op &
shutdown
Fixed Display Data Query in INVERT format & printout showing non- invert
Measurment Results Duplicate entry in Setup & Info
Fixed remote command :DISPlay:LABel causes DCA to freeze in some remote
sequences
Can't format floppy drive when full of waveform files
Signal goes to the bottom of screen at various vertical scale settings i.e. 6.1mV/div
Fixed remote command :WAV:FORM WORD;SOURC CGR;DATA?
Measurement Enhancements
Improved built in Jitter measurement. The algorithm now uses a smaller window (3 pixels high)
about the crossing point of the eye.
The EYE measurement is based on the histogram database.
In the case of jitter measurement, we need a "window" with vertical boundares of [y1,y2] and
horizontal [x1,x2]. The jitter value is simply derived from the histogram within this window, e.g.
the jitter RMS is the standard deviation of the histogram. The smaller the window, the smaller
the jitter.
In firmware versions previous to A.02.10, 5% of the [Ymax,Ymin] is used, assuming that the
"crossing" position is Ycross,
then

Table of Contents

Other manuals for Agilent Technologies 86100C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 86100C and is the answer not in the manual?

Agilent Technologies 86100C Specifications

General IconGeneral
BrandAgilent Technologies
Model86100C
CategoryTest Equipment
LanguageEnglish

Related product manuals