Chapter 1 1-27
Alphabetical Command Reference
CALK
CALK
Syntax
CALK<24MM|292MM|292S|32F|35MC|35MD|35ME|716|7MM|N50|N75|TRLK|USED>; or
CALK<24MM|292MM|292S|32F|35MC|35MD|35ME|716|7MM|N50|N75|TRLK|USED>?;
Description
Command Description Range Query Response
CALK24MM Selects a 2.4-mm calibration kit
(85056A/D) as the default cal kit.
N/A
<0|1><
L
F
>
CALK292MM Selects a 2.92-mm calibration kit as the
default cal kit.
N/A
<0|1><
L
F
>
CALK292S Selects a 2.92* calibration kit (85056K) as
the default cal kit.
N/A
<0|1><
L
F
>
CALK32F Selects a 32F calibration kit (85032F) as
the default cal kit.
N/A
<0|1><
L
F
>
CALK35MC
1
1. Use
CALK35MM for the 8752C and 8753D analyzers to select the 85033C cal kit.
Selects a 3.5-mm calibration kit (85033C)
as the default cal kit.
N/A
<0|1><
L
F
>
CALK35MD Selects a 3.5-mmcalibration kit (85052for
8720ET/ES series analyzers, and
85033D/E for 8753ET/ES analyzers) as
the default cal kit.
N/A
<0|1><
L
F
>
CALK35ME Selects a 33D/E calibration kit (85033D/E)
as the default cal kit for all analyzers.
N/A
<0|1><
L
F
>
CALK716 Selects a 7-16 calibration kit (85038) as
the default cal kit.
N/A
<0|1><
L
F
>
CALK7MM Selects a 7-mm calibration kit (85050
series for8720ET/ES series analyzers,and
85031B for 8753ET/ES analyzers) as the
default cal kit.
N/A
<0|1><
L
F
>
CALKN50 Selects a type-N 50 ohm calibration kit
(85054 for 8720ET/ES series analyzers,
and 85032B/E for 8753ET/ES analyzers)
as the default cal kit.
N/A
<0|1><
L
F
>
CALKN75 Selects a type-N 75 ohm calibration kit
(85036B/E) as the default cal kit.
N/A
<0|1><
L
F
>
CALKTRLK Selects a TRL 3.5-mm calibration kit
(85052C) as the default cal kit.
N/A
<0|1><
L
F
>
CALKUSED Selects a user-defined calibration kit. N/A
<0|1><
L
F
>