EasyManua.ls Logo

Agilent Technologies 8960 Series - Change the MS TX Level During a Handover

Agilent Technologies 8960 Series
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
66 Chapter 3
GPRS Mobile Test Application
How Do I Change the MS TX Level?
B. Change the MS TX level during a handover.
1. Press the CALL SETUP key.
2. On Control menu, press the Handover Setup (
F5) key.
3. Select and change the MS TX Level.
4. Press the Handover Execute (
F5) key to change the MS TX level, or press the Close
Menu (
F6) key to leave the level unchanged.
5. Use the
More key to check that the MS TX level has been changed on the Call Parms menu
3 of 3.
3
4
1
5
深圳德标仪器 135-1095-0799

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Related product manuals