EasyManuals Logo
Home>Agilent Technologies>Power Supply>E3641A

Agilent Technologies E3641A User Manual

Agilent Technologies E3641A
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
81
4
VOLTage:PROTection? [MINimum | MAXimum]
Query the overvoltage protection trip level presently programmed.
VOLTage:PROTection:STATe {0 | 1 | OFF | ON}
Enable or disable the overvoltage protection function. At
*RST
, this value is
set to ‘‘ON’’.
VOLTage:PROTection:STATe?
Query the state of the overvoltage protection function. The returned parameter
is ‘0’’ (OFF) or ‘‘1’(ON).
VOLTage:PROTection:TRIPped?
Return a ‘1 if the overvoltage protection circuit is tripped and not cleared or
a ‘0’’ if not tripped.
VOLTage:PROTection:CLEar
Cause the overvoltage protection circuit to be cleared. After this command,
the output voltage is restored to the state it was in before the protection feature
occurred and the OVP trip level remains unchanged to the value presently
programmed. Before sending this command, lower the output voltage below
the trip OVP point, or raise the OVP trip level above the output setting. Note
that the overvoltage condition caused by an external source must be removed
first before proceeding this command.
VOLTage:RANGe {P8V* | P20V* | P35V** | P60V** | LOW | HIGH}
Select an output range to be programmed by the identifier. For example,
‘‘P20V’’ or ‘HIGH’is the identifier for the 20V/1.5A range and P8V’’ or ‘LOW’’
is for the 8V/3A range (for E3640A model). At
*RST
, low voltage range is
selected.
VOLTage:RANGe?
Query the currently selected range. The returned parameter is P8V or ‘P35V
for low voltage range, or ‘‘P20V’’ or ‘‘P60V for high voltage range.
MEASure:CURRent?
Query the current measured across the current sense resistor inside the power
supply.
MEASure[:VOLTage]?
Query the voltage measured at the sense terminals of the power supply.
* For E3640A/42A/44A models **For E3641A/43A/45A models

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3641A and is the answer not in the manual?

Agilent Technologies E3641A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3641A
CategoryPower Supply
LanguageEnglish

Related product manuals