Chapter 4 Remote Interface Reference
Triggering
83
4
To ensure synchronization when the bus source is selected, send the
*WAI
(wait) command. When the
*WAI
command is executed, the power supply
waits for all pending operations to complete before executing any additional
commands. For example, the following command string guarantees that the
first trigger is accepted and is executed before the second trigger is
recognized.
‘‘TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
’’
You can use the
*OPC?
(operation complete query) command or the
*OPC
(operation complete) command to signal when the operation is complete.
The
*OPC?
command returns ‘‘1’to the output buffer when the operation
is complete. The
*OPC
command sets the OPC bit (bit 0) in the Standard
Event register when the operation is complete.
Immediate Triggering
To select the immediate trigger source, send the following command.
‘‘TRIG:SOUR IMM
’’
When the
IMMediate
is selected as a trigger source, an
INITiate
command immediately transfers the
VOLT:TRIG
or
CURR:TRIG
value to
VOLT
or
CURR
value. Any delay is ignored.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3641A and is the answer not in the manual?

Agilent Technologies E3641A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3641A
CategoryPower Supply
LanguageEnglish

Related product manuals