Installation Guide 17
E440xBU Option H70 Safety and Regulatory Information
Servicing
Warnings applicable to this instrument
.
WARNING To prevent electrical shock, disconnect the instrument from the mains
before cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
WARNING For continued protection against fire hazard replace line fuse only with
same type and rating: Fuse 3.15A/250V, Part Number 2110-0655
The use of other fuses or material is prohibited.
WARNING This is a Safety Class I product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall be inserted only into
a socket outlet provided with a protective earth contact. Any interruption
of the protective conductor, inside or outside the product is likely to make
the product dangerous. Intentional interruption is prohibited.
WARNING These servicing instructions are for use by qualified personnel only.
WARNING The opening of covers or removal of parts is likely to expose dangerous
voltages. Disconnect the instrument from all voltage sources while it is
being opened.
WARNING This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 61010-1: 2001.
WARNING No operator serviceable parts inside. Refer servicing to qualified personnel.
WARNING If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.