4.4.6.2.2.2 Ref Value
Allows you to set the absolute power reference.
4.4.6.2.2.3 Ref Position
Allows you to set the display reference position to either Top, Ctr (center), or Bot (Bottom).
Example: :DISP:PMD:WIND:TRAC:Y:PDIV 20
:DISP:PMD:WIND:TRAC:Y:PDIV?
Key Path:
Amplitude Y Scale
Remote Command: :DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:RLEVel
<angle>
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:RLEVel
?
Unit: rad
Preset: 0.00
State Saved: Saved in instrument state.
Min: – 628 rad
Max: 628 rad
Dependencies/
Couplings:
See Restriction and Notes
Notes: If the Scale Coupling is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Scale Coupling automatically
changes to Off.
Example: :DISP:PMD:WIND:TRAC:Y:RLEV 0
:DISP:PMD:WIND:TRAC:Y:RLEV?
Key Path:
Amplitude Y Scale
Remote Command: :DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:RPOSition
TOP|CENTer|BOTTom
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:Y[:SCALe]:RPOSition
?
Preset: CENTer
State Saved: Saved in instrument state.
Range: Top | Ctr | Bottom