Chapter 2 157
Front-Panel Key Reference
MEASURE
quasi-peak detector measurement for
the specified dwell time. Default is On
with a 200 ms dwell time.
Key Access:
MEASURE, More,
Auto Measure
Average
On Off
Allows you to turn average
measurements On and Off. When
Average is On, you can change the
measurement dwell time. If this
detector is On, the Measure at Marker
and Auto-measure functions make an
average detector measurement for the
specified dwell time. Default is On with
a 200 ms dwell time.
Key Access:
MEASURE, More, Signal List,
Remeasure, MeasDetector
Range Dwell
Specifies the amount of time the
analyzer dwells on a signal during
ranging. Ranging is the first step in a
signal measurement. During ranging,
the analyzer sweeps for the specified
range dwell time, before adjusting the
signal peak to the top of the screen for
further measurements. If system
defaults are restored, Range Dwell will
default to 200 ms. See “Dwell Time
Settings” in the Measurement
Techniques” Chapter of the Agilent
EMC Analyzer Measurement Guide for
more information.
Key Access:
MEASURE, More, Signal List,
Remeasure, RemeasDetector
Center Presel
On Off
Agilent E7403A, E7404A, and E7405A
only. Allows you to turn the center
preselector filter On or Off. This
determines whether or not you do a
preselector center before each
measurement in the preselected bands.