Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 97
4 Preparing the Module
Install Capillary Connections
For correct installation of capillary connections it's important to choose the
correct fittings, see “Syntax for Capillary Description” on page 92.
Table 18 Capillary connections for 1290 Infinity II Bio LC
p/n From To
Bottle Head Assembly (G7120-60007) Solvent Bottle Infinity II Pump
Capillary MP35N 0.17 mm x 500 mm, SI/SI (5500-1419) Pump Multisampler
Capillary MP35N 0.12 mm x 500 mm SI/SI (5500-1279) Multisampler MCT
Quick Connect Capillary MP35N 0.12 mm x 105 mm
(5500-1578)
MCT Heat Exchanger Column
Quick Turn Capillary MP35N 0.12 mm x 280 mm (5500-1596) Column/MCT Valve Detector
Waste accessory kit (5062-8535) VWD Waste
Tube PTFE 0.7 mm x 5 m, 1.6 mm od (5062-2462) DAD/FLD Waste
Analytical tubing kit 0.25 mm i.d. PTFE-ESD (G5664-68712) Detector Fraction Collector